2150 mm390–800 mm

Acute and high-dependency beds

VR-5 acute care bed

Four-section bed for high-dependency and acute wards, with integrated scale, CPR release and a 390 mm low position.

Safe working load
230 kg
Platform height
390 to 800 mm
Backrest
0 to 70°
Battery
24 V lithium-ion

Ask for a demonstration Documents

Overview

The VR-5 is Van Roon’s bed for high-dependency units and acute wards. The current VR-5, introduced in 2016, uses the same controller family as the VR-7, so the two beds share most of their spare parts, their service tool and their training.

It goes lower than the VR-7, to 390 mm, for patients at risk of falls, and has the same mechanical CPR release, integrated scale and patient-exit detection. It does not have lateral tilt.

  • Low position at 390 mm

    With a 150 mm mattress, the patient sits with their feet on the floor when getting in and out of bed.

  • Chair position

    The backrest and leg section move together to a seated position, held by one button on the nurse handset.

  • CPR release

    Red handles on both sides bring the backrest flat without power.

  • Trendelenburg

    12° head down from the nurse handset, and by a dedicated button on the foot-end control panel.

  • Integrated scale

    Four load cells, 0.1 kg resolution, shown on the foot-end display.

  • Patient-exit detection

    Two zones, with an alarm at the bed and a relay to the ward’s nurse call system.

Specifications

Dimensions
Overall length2,150 mm (2,350 mm with the length extension)
Overall width1,020 mm side rails raised
Mattress platform2,000 × 860 mm
Platform height (without mattress)390 to 800 mm
Under-bed clearance150 mm
Weight of bed146 kg without mattress
Positions
Backrest0 to 70°
Thigh section0 to 30°
Trendelenburg and reverse Trendelenburg12° each way
One-touch positionsFlat, chair, CPR
Loads
Safe working load230 kg
Maximum patient weight170 kg
Weighing and patient-exit detection
ScaleIntegrated; resolution 0.1 kg; accuracy ±0.5 kg from 20 to 230 kg
Patient-exit detectionTwo zones; nurse call relay
Electrical
Supply230 V AC, 50 Hz; maximum consumption 400 VA
ProtectionClass I; type B applied part; IPX4
Battery24 V lithium-ion pack (from 2025); earlier beds 2 × 12 V lead-acid
Mobility
Castors125 mm, central braking and steering
Regulatory and service
ClassificationMedical device, Class I; CE and UKCA marked
StandardsIEC 60601-1, IEC 60601-1-2, IEC 60601-2-52
Expected service life10 years
Planned maintenanceEvery 12 months
ManufacturerVan Roon Medische Techniek B.V., Leiden, the Netherlands

Dimensions are nominal and measured without a mattress unless stated. Safe working load is the total load the bed is designed to carry, including the patient, the mattress and any accessories fitted. Specifications may change without notice; the instructions for use supplied with each bed take precedence.

Options and accessories

Part numberDescriptionNotes
VR5-0410IV pole, two-stageMaximum load 12 kg
VR5-0350Nurse handsetReplaces the handset supplied with the bed
VR5-0352Patient handset with lockoutFunctions can be locked from the foot-end panel
VR5-0445Patient helperMaximum load 75 kg on the handle
VR5-0480Remote diagnostics moduleAs VR-7
VR5-0590Battery pack, 24 V lithium-ionReplaces the lead-acid pack VR5-0588; controller firmware 4.10 or later

Options marked as retro-fit kits are fitted by a Van Roon engineer or by an engineer holding a current Level 2 certificate. For prices, contact sales@vanroon.uk. Spare parts are on the spare parts page.

Surfaces

The VR-5 is supplied with the mattress chosen for the unit. These Van Roon surfaces are tested with it:

Mattresses from other manufacturers may change the effective height of the side rails. Check the rail height above the compressed mattress against the instructions for use before using any other mattress.

Documents

DocumentTypeNumberRevisionDate
VR-5 acute care bed: instructions for use
PDF, 142 KB, 22 pages
Instructions for use VRM-IFU-0501-EN-GB L
Cleaning and decontamination of VR-series beds and surfaces
PDF, 29 KB, 2 pages
Guide VRM-DG-0005-UK F
VR-5: declarations of conformity
PDF, 25 KB, 2 pages
Declaration of conformity VRM-DOC-0501 9

Printed instructions for use are supplied with every bed. If you need a printed copy, or a copy in another format, contact info@vanroon.uk. Service documentation is available to trained engineers only.

Service

The VR-5 needs planned maintenance every 12 months, including an electrical safety test, a check of every movement and alarm, a load test of the side rails and a calibration check of the scale. Van Roon engineers carry the parts most often needed, so most faults are fixed at the first visit.

Hospital engineers who hold a current Level 2 certificate for the VR-5 can carry out planned maintenance and most repairs themselves. See technical training and service agreements.

Customer notices for the VR-5

Service desk 0808 157 0614

VR-5 is manufactured by Van Roon Medische Techniek B.V., Leiden, the Netherlands, and supplied in the United Kingdom and Ireland by Van Roon Medical UK Ltd. Available in the UK and Ireland since 2004. Always read the instructions for use before using the product.