Each VR-series bed runs a software package made up of the controller firmware, which moves the bed and runs its safety functions, and the panel software, which drives the footboard panel and the side rail controls.
Notes
Controller firmware 4.12 installed on all beds
The update announced in CI-UK-2025-06 is complete. Every VR-series bed on our service records in the UK and Ireland runs controller firmware 4.12.
VR-7 2025.2 and VR-5 2025.3 released
Both packages include controller firmware 4.12. Van Roon engineers will install them at each bed’s next planned maintenance visit. See customer information CI-UK-2025-06.
Current releases in the UK and Ireland
| Product | Package | Controller firmware | Panel software | Released |
|---|---|---|---|---|
| VR-7 and VR-7 XW | 2025.2 | 4.12 | 3.8.1 | |
| VR-5 | 2025.3 | 4.12 | 2.6.4 | |
| VR-A4 pump | 1.9 | Not applicable | 1.9.0 |
The package number is shown on the footboard panel under Settings › About this bed. The VR-A4 pump shows its version for three seconds when it is switched on.
How updates are installed
Software is not available to download from this website. Packages are installed with the Van Roon service tool, a Windows program that connects to the bed’s USB service port, by Van Roon engineers or by hospital engineers who hold a current Level 2 certificate for the product.
Where an update is needed for safety, we write to every hospital concerned with a customer notice and install it at no charge. Other updates are installed at the next planned maintenance visit for beds under a Van Roon service agreement, and are available to Level 2 engineers on request.
Beds fitted with the remote diagnostics module (VR7-0480 or VR5-0480) send their event log to the Van Roon service centre in Leiden when the hospital allows it. Software cannot be installed through the remote diagnostics module.
To ask for a package, give the serial numbers of the beds and the name of the Level 2 engineer who will install it to the service desk on 0808 157 0614 or service@vanroon.uk.
Release history
| Package | Released in the UK | Controller | Panel | Changes |
|---|---|---|---|---|
| VR-7 2025.2 | 4.12 | 3.8.1 | Controller: stronger server identity checks in the remote diagnostics interface (CI-UK-2025-06). Panel: corrected Welsh-language strings on the weighing screen. | |
| VR-5 2025.3 | 4.12 | 2.6.4 | Controller as VR-7 2025.2. | |
| VR-7 2025.1 | 4.10 | 3.8.0 | Controller: charge profile for 24 V lithium-ion battery packs (CI-UK-2025-03). Panel: charge profile setting; battery health screen. | |
| VR-5 2025.1 | 4.10 | 2.6.2 | Controller as VR-7 2025.1. Panel: charge profile setting. | |
| VR-7 XW 2024.1 | 4.06 | 3.7.2 | First release for the VR-7 XW. | |
| VR-A4 pump 1.9 | – | 1.9.0 | Quieter alternation at night setting. Corrected the low-pressure alarm delay to 90 seconds. | |
| VR-7 2024.1 | 4.06 | 3.7.2 | Controller: smoother start and stop of lateral tilt. Panel: weight trend extended to 30 readings; Irish-language option. | |
| VR-5 2023.2 | 4.04 | 2.5.9 | Controller: patient-exit detection zone 1 made less sensitive to patient turning. | |
| VR-7 2023.1 | 4.04 | 3.6.0 | Controller as VR-5 2023.2. Panel: VR-A4 surface control from the footboard. |
Release dates are the dates each package was approved for installation in the UK and Ireland. Earlier releases are listed in the service documentation.